Agilent Technologies 1670G Printer Accessories User Manual


 
MODE
Command :TRIGger:MODE {EDGE|PATTern|IMMediate}
The MODE command allows you to select the trigger mode for the
oscilloscope. In the IMMediate trigger mode, the oscilloscope goes to a
freerun mode and does not wait for a trigger. Generally, the IMMediate mode
is used when correlating measurements with the analyzer.
In EDGE trigger mode, the oscilloscope triggers on an edge of a waveform,
specified by the SOURce, DELay, LEVel, and SLOPe commands. If a source
is not specified, then the current source is assumed.
In PATTern trigger mode, the oscilloscope triggers when entering or exiting a
specified pattern of the two internal channels and external trigger. The
pattern is generated using the CONDition, DELay, LEVel, LOGic and PATH
commands. The CONDition command allows the oscilloscope to trigger when
entering the specified pattern or exiting the pattern. The DELay value
corresponds to the Count field displayed on the TRIGger menu. The LOGic
command defines the pattern. The PATH command is used to change the
trigger pattern and level. The path consists of two channels.
Example OUTPUT XXX;":TRIGGER:MODE PATTERN"
Query :TRIGger:MODE?
The MODE query returns the current trigger mode selection.
Returned Format
[:TRIGger:MODE] {EDGE|PATTern|IMMediate}<NL>
Example OUTPUT XXX;":TRIGGER:MODE?"
TRIGger Subsystem
MODE
35-11