XAUTo
Command :MARKer:XAUTo {MANual|CHANnel<N>,<type>,<level>,
<slope>,<occurrence>}
The XAUTo command specifies the automatic placement specification for
the X marker. The first parameter specifies if automarker placement is to be
in the Manual mode or on a specified channel. If a channel is specified, four
other parameters must be included in the command syntax. The four
parameters are marker type, level, slope, and the occurrence count.
<N> {1|2}
<type> {ABSolute | PERCent}
<level> percentage of waveform voltage level, ranging from 10 to 90 of the Vtop to
Vbase voltage or a voltage level
<slope> {POSitive | NEGative}
<occurrence> integer from 1 to 100
Example OUTPUT XXX;":MARKER:XAUTO CHANNEL1,ABS,4.75,POSITIVE,5"
Query :MARKer:XAUTo?
The XAUTo query returns the current settings.
Returned Format
[:MARKer:XAUTo] {MANual | CHANnel<N>,<type>,
<level>,<slope>,<occurrence>}<NL>
Example OUTPUT XXX;":MARKER:XAUTO?"
MARKer Subsystem