Agilent Technologies 1670G Printer Accessories User Manual


 
TINTerval:AUTorange
Command :SPA{1|2}:TINTerval:AUTorange
{LOGarithmic|LINear},<min_time>,<max_time>
The TINTerval:AUTorange command automatically sets the Time Interval
ranges in a logarithmic or linear distribution over the specified range of time.
When the AUTorange command is executed, the data accumulators and
statistic counters are reset.
<min_time> real number
<max_time> real number
Example
OUTPUT XXX;":SPA2:TINTerval:AUTorange LINear,4.0E-3,55.6E+2"
OUTPUT XXX;":SPA1:TINTerval:AUTorange LOGarithmic,3.3E+1,8.6E+2"
TINTerval:QUALifier
Command :SPA{1|2}:TINTerval:QUALifier
<label_name>,<start_pattern>,<end_pattern>
The TINTerval:QUALifier command defines the start and stop patterns for a
specified label. The start and stop patterns determine the time windows for
collecting data. A query returns the currently defined start and stop patterns
for a given label.
<label_name> string of up to 6 alphanumeric characters
<start_pattern> "{#B{0|1}...|
#Q{0|1|2|3|4|5|6|7}...|
#H{0|1|2|3|4|5|6|7|8|9|A|B|C|D|E|F}.. .|
{0|1|2|3|4|5|6|7|8|9}...}"
SPA Subsystem
TINTerval:AUTorange
25–22