Kodak 5500 Scanner User Manual


 
5-16 A-61070 September 1999
7. Carefully insert the green rear lamp sock holder and rear lamp into
the scanner frame. Be sure to insert the lamp socket holder straight
in along its axis.
8. Connect the rear lamp cable connector to the scanner electrical
system.
9. Repeat Steps 3 through 8, referring to the front lamp, noting that in
Step 3 the front lamp cable must be released from the front lamp
cable clip, and in Step 6 the clear slit should face towards the rear.
In Step 8, the front lamp cable connector must be fashioned to the
front lamp cable clip.
10. Close the side access door.
11. Perform the Calibration function. See the section entitled,
“Calibrating the scanner” in Chapter 2.