Xantrex Technology XPR 20-300 Power Supply User Manual


 
Operation
Using Over Voltage Protection (OVP)
Release 1.1 51
Using Over Voltage Protection (OVP)
The OVP circuit protects the load in the event of a remote programming error, an
incorrect voltage control adjustment, or a power supply failure. The protection
circuit monitors the output voltage at the output of the power supply and will shut
down the main power converter whenever a preset voltage limit is exceeded. Set the
preset voltage limit (also called the set point or trip level) using the
screwdriver-adjustable, front panel OVP potentiometer (OVP set).
Front Panel
OVP
Operation
In local control mode, check the OVP set point at any time by pressing the OVP
CHECK switch. The OVP set point is the value displayed on the digital voltmeter.
To set the trip level from the front panel:
1. Disconnect any loads. Turn the power supply ON.
2. Adjust the power supply output voltage to any voltage lower than the desired trip
level.
3. Press the front panel OUT ON/OFF (output shutdown) switch to its IN position.
The OFF LED turns on.
4. Press the OVP switch to see the OVP set point on the voltmeter display.
5. Holding down the OVP switch, turn the OVP SET potentiometer until the
desired set point is reached. Release the OVP switch.
6. Press the OUT ON/OFF switch to its OUT position. The ON LED turns on.
To check that the power supply shuts off at the desired set point, slowly increase the
output voltage while monitoring the front panel voltmeter. The OVP LED on the
front panel turns on when the OVP circuit activates, and the output voltage will
quickly fall to zero.
Note The default preset limit is approximately 110% of the rated output voltage.
When using OVP with remote sensing lines connected, compensate for the voltage line
drop across the output return line by measuring or calculating the line drop, then
adding this value to the desired OVP setpoint.