Xantrex Technology XPR 20-300 Power Supply User Manual


 
Specifications and Characteristics
Output Performance Specifications
72 Operating Manual for XPR Series Power Supply
Programming Accuracy
1
Readback Accuracy
30 Minute Drift
2
8 Hour Drift Temperature Stability
3
Temperature Coefficients
Remote Analog Programming Interface
Voltage Programming
Resistive Programming
1% of V or Imax
2% of V or Imax
1. Accuracy specifications apply for settings in range of 1% to 100% of rated output
Front Panel or Remote Analog Programming Interface
Voltage Readback
Current Readback
1% of Vmax
1% of Imax
Voltage 0.04% of Vmax
Current 0.6% of Imax
2. At 25°C ±5°C, with full power load
Voltage 0.02% of Vmax
Current 0.04% of Imax
3. At 25°C ±5°C after 30 minutes full load operation
Front Panel or Analog Programming Interface
Voltage Programming
Current Programming
Voltage Readback
Current Readback
0.04% of Vmax
/°C
0.06% of Imax/°C
0.04% of Vmax/°C
0.06% of Imax/°C