Agilent Technologies 66001A Mouse User Manual


 
12 General Information
Table 1-1. Performance Specifications (continued)
Agilent Model
Parameter 66101A 66102A 66103A 66104A 66105A 66106A
Line Regulation (change in output
voltage or current for any line change
within ratings)
Voltage: 0.5 mV 0.5 mV 1 mV 2 mV 3 mV 5 mV
Current: 0.75 mA 0.5. mA 0.3 mA 0.1 mA
50
µ
A 30
µ
A
Transient Response Time (for output
voltage to recover to within 100 mV of
its previous level following a step
change in load current that is up to 10%
of the power module rated output
current.
< 1 ms
AC Input Ratings (selectable via
module switch- see Chapter 2)
Nominal line voltage (Vac)
115 V switch setting 87- 132 Vac
230 V switch setting 174- 250 Vac
Frequency range: 47 - 63 Hz
Output Terminal Isolation (maximum,
from output to the mainframe chassis
ground):
±240 Vdc
Table 1-2. Supplemental Characteristics
Agilent Model
Parameter 66101A 66102A 66103A 66104A 66105A 66106A
Output Programming Range
(Maximum programmable values):
Voltage: 8.190 V 20.475 V 35.831 V 61.425 V 122.85 V 204.75 V
Current: 16.380 A 7.678 A 4.607 A 2.559 A 1.280 A 0.768 A
OV Protection: 9.6 V 24.0 V 42.0 V 72.0 V 144.0 V 240.0 V
Average Programming Resolution
Voltage: 2.4 mV 5.9 mV 10.4 mV 18.0 mV 36.0 mV 60.0 mV
Current: 4.6 mA 2.3 mA 1.4 mA 0.75 mA 0.39 mA 0.23 mA
OV Protection: 50 mV 120 mV 200 mV 375 mV 750 mV 1.25 V
OVP Accuracy (@ calibration
temperature ±5
°
C):*
250 mV 500 mV 800 mV 1 V 1.5 V 2.5 V
* Factory calibration temp = 25
°
C
Readback Resolution
Voltage: 0.305 mV 0.763 mV 1.335 mV 2.289 mV 4.577 mV 7.629 mV
Current: 0.587 mA 0.293 mA 0.169 mA 0.099 mA 0.049 mA 0.030 mA
Drift (Temperature Stability)
(following a 30-minute warmup, change
in output over 8 hours under constant
line, load, and ambient temperature)
Voltage: 0.02% + 0.4 mV 1 mV 2 mV 3 mV 6 mV 10 mV
Current: 0.02% + 8 mA 5 mA 2.5 mA 1.3 mA 0.63 mA 0.38 mA