Table 1-2. Supplemental Characteristics (continued)
Agilent Model
Parameter 66101A 66102A 66103A 66104A 66105A 66106A
Temperature Coefficients (change in
output per
°
C after a 30-minute
warmup)
Output Voltage: 30 ppm + 0.1 mV 0.2 mV 0.3 mV 0.6 mV 1.1 mV 1.8 mV
Output Current: 30 ppm + 0.2 mA 0.1 mA 0.05 mA 0.03 mA 0.01 mA 0.006 mA
Voltage Readback: 20 ppm + 0.1 mV 0.2 mV 0.2 mV 0.7 mV 0.7 mV 1.1 mV
±
Current Readback:
20 ppm + 0.2 mA 0.1 mA 0.05 mA 0.02 mA 0.02 mA 0.02 mA
OV Protection: 1 mV 2 mV 4 mV 8 mV 16 mV 30 mV
Maximum Reverse Current
(maximum current, without damage,
that the power module will withstand
while turned on and with the dc output
reverse biased by an external dc
source):
16 A 7.5 A 4.5 A 2.5 A 1.25 A 0.75 A
Typical Output Common Mode Noise
Current*
rms: 0.5 mA
peak-peak: 5 mA
* Referenced to mainframe chassis ground.
Remote Sensing Capability
Voltage drop per lead: Up to 1/2 of rated output voltage
Load voltage: Subtract voltage drop in load leads from specified output voltage rating
Load Regulation: For each 1-volt change in the negative output lead due to load current change,
add 2 mV to the voltage load regulation specification (see Table 1-1).
Command Processing Time (average
time for output to begin to change after
receipt of digital data when the power
module is connected directly to the
GPIB):
20 ms
Output Voltage Programming
Response Time*
Time for output to change from 10% to
90% of its total excursion: < 20 ms
Time for output to change from 90% to
10% of its total excursion:
< 20 ms < 50 ms
Time for output voltage to settle to
within 0.1% of the final value: 120 ms