Agilent Technologies N9340A Network Card User Manual


 
SCPI Command Reference 6
N9340A User’s Guide 117
INITiate Subsystem
The INITiate subsystem is used to control the
initiation of the trigger. Refer to the TRIGger
subsystem for related commands.
Continuous or Single Sweep
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Toggles the continuous sweep between on and off.
When the On (1) parameter is selected, the
continious sweep is enabled.
When the Off (0) parameter is selected, the
continuous sweep is disabled and single sweep is
enabled.
Initiate a single sweep
:INITiate[:IMMediate]
This command initiates a single sweep.
Note that the command becomes effective only
after you set the sweep mode to single sweep
previously.
*RST State: Continuous sweep (On)
Key Access: [BW/SWP] > {Sweep}
Key Access: [BW/SWP] > {Single Sweep}