Agilent Technologies N9340A Network Card User Manual


 
Key Reference 5
N9340A User’s Guide 73
BW/SWP
Activates the bandwidth function and accesses the
associated softkeys to control resolution bandwidth,
video bandwidth and sweep time.
RBW
Changes the 3-dB resolution bandwidth on the
analyzer from 30 Hz to 1 MHz in a 1, 3, 10
sequence using the knob, step keys or the numeric
keypad.
As the resolution bandwidth is decreased, the
sweep time is modified to maintain amplitude
calibration. Resolution bandwidth is also related to
span. As span is decreased, resolution bandwidth
decreases. As the resolution bandwidth changes,
the video bandwidth, if in auto couple mode,
changes to maintain the VBW/RBW ratio.
A “#”mark appears next to RBW on the display
when it is not coupled.
Key access: [BW/SWP]
NOTE
If an unavailable bandwidth is entered using the numeric
keypad, the closest available bandwidth in the 1, 3, 10
sequence is used.
NOTE
In non-zero span mode, the value of RBW is determined by
whether the setting for RBW is Auto or Man (manual).