Agilent Technologies N9340A Network Card User Manual


 
SCPI Command Reference 6
N9340A User’s Guide 127
[:SENSe]:DETector Subsection
Automatic Detection Type Selected
[:SENSe]:DETector:TRACe[1]|2|3|4 AUTO
OFF|ON|0|1
[:SENSe]:DETector:TRACe[1]|2|3|4 AUTO?
This command switches automatically to the
optimum detection type for typical measurements
using the current instrument settings.
The detector type is average if any of these are on:
Noise marker
Band power markers
Trace averaging when the Average Type is Power
(RMS).
The detector type is sample if any of the following
conditions are true:
Trace averaging is on with average type of video
Both max. and min. hold trace modes are on
Resolution bandwidth is less than 1 kHz, and
noise marker, band power markers, or trace
averaging is on.
The detector type is negative peak if any trace is
in minimum hold and no traces are in max hold.
The detector type is peak if the above conditions
are off.
Manually changing the detector function turns Auto
off.
*RST State: On
Key
Access: [TRACE] > {More (1 of 2)} > {Detector}