Agilent Technologies N9340A Network Card User Manual


 
Getting Started 2
N9340A User’s Guide 21
Viewing a Signal
1 Use a signal generator to generate a
continuous- wave signal (1 GHz, 0 dBm).
2 Press [SYS] > {More (1 0f 3)} > {Preset} and select
Default to toggle the preset setting to
factory- defined status.
3 Press the green [Preset] key to restore the
analyzer to its factory-defined setting.
4 Connect the generator’s RF OUT connector to
analyzer’s RF IN connector on the top panel.
5 Press [FREQ] > {Center Freq} > 1> {GHz} to set the
center frequency to 1 GHz.
6 Press [SPAN] > 5 > {MHz} to set the frequency
span to 5 MHz.
7 Press [MARKER] > {Peak Search} > {Peak} to place a
marker (labeled 1) at the highest peak (1 GHz)
on the display.
The values of amplitude and frequency of the
Marker appear both in the function block and in
the upper-right corner of the screen.
Use the knob, the arrow keys, or the softkeys in
the Peak Search menu to move the marker and
read out the value of both frequency and
amplitude displayed on the screen.
Figure 1 View a signal (1 GHz, 0 dBm)