Emerson Process Management 3-9000-522 Whiteboard Accessories User Manual


 
APPLICATION FUNCTIONS 5-3
MON2000
JULY 2010 System
Table 5-1 Description of Items in System Dialog
Item Function Edit
Stream Sequence Defines the order of stream analysis for Detector 1.
Stream Sequence 2 Defines the order of stream analysis for Detector 2;
only available with selected hardware.
Note that this item is not available in all GC applica-
tions.
Aux. Stream Sequence Defines the auxiliary order of stream analysis for
Detector 1. Can be used to run an alternative stream
sequence when a Discrete Input is set (see Section 5.11
for detailed instructions).
Note that this item is not available in all GC applica-
tions.
Aux. Stream Sequence 2 Additional alternative stream sequence provided for
second detector in dual detector applications.
Auxiliary order of stream analysis for Detector 2. Can
be used to run an alternative stream sequence when a
Discrete Input is set (see Section 5.11 for detailed
instructions).
Note that this item is not available in all GC applica-
tions.
Analyzer Name Defines the GC unit name that appears in the status
bar of the main window when connected to the GC Con-
troller. Can contain up to 12 characters.
Unit Type Identifies GC unit type (i.e., model number).
System Description A field to record miscellaneous reference
information to further identify the system. Can contain
up to 28 characters.
Chromatogram Buffer Size Provides number of bytes used to store a
chromatogram.
Max Peaks Provides maximum number of peaks (application-spe-
cific).