Emerson Process Management 3-9000-522 Whiteboard Accessories User Manual


 
5-4 APPLICATION FUNCTIONS
MON2000
System JULY 2010
Number of Valves Identifies number of chromatographic valves
(application-specific).
Daylight Savings Time Allows the GC Controller to adjust automatically for
DST. “Normal” enables automatic DST adjustment
(United States). “Override” disables automatic DST
adjustment.
CGM Analog Output Displays chromatogram trace signals for strip chart
generation from Detectors 1 and/or 2, trace 1 only.
Identifies the analog output number.
If this field is set to 0, then no analog output will
produce the CGM signal.
CGM Baseline Offset A fixed offset (percent of full scale) added to the CGM
analog output to allow the user to conveniently position
the chromatogram on a strip chart recorder. Displays
chromatogram trace signals for strip chart generation
from Detectors 1 and/or 2. Identifies the baseline offset.
CFG Baseline Number Identifies the Standard GC Application software from
which this GC application or configuration was
developed (i.e., the number of the standard configura-
tion file that was used to generate the current applica-
tion).
Metric Base Condition Enables (metric) or disables (off) whether MON2000
displays the GC Control Calculation in metric units.
See Section 5.5.1 for details.
Table 5-1 Description of Items in System Dialog (Continued)
Item Function Edit
When assigning a CGM Analog Output, also check the
Analog Outputs dialog (see Section 5.10). A CGM
Analog Output setting that uses a previously assigned
analog output number will produce erratic CGM trace
output, possibly with incorrect scaling.
Do not assign this item analog output number 1 if the
connected GC unit supports bargraph output
(see Section 5.10.3 for details).