Tektronix 070-9180-01 Computer Hardware User Manual


 
Command Syntax
Option 01 VXI Interface Module User Manual
3–9
For commands directed to a VX4372 or VX4374 which has two scanner
sections:
<section_list> Sections Specified
(1) Section 1
(1,2) Sections 1 and 2
(1:2) Sections 1 and 2
error/event queue When the command parser detects a syntax error or data range
error, it places an error message describing the error in the Error/Event queue.
Bit 2 of the Status Byte Register is set to indicate that this queue is not empty.
Bit 5 of the Standard Event Status Register (the Command Error bit) is set if the
parser detects a syntax error. Bit 4 of the Standard Event Status register (the
Execution Error bit) is set if the parser detects a numeric argument that is out of
range. When a SYSTem:ERRor? query is received, an error message is dequeued
from the Error/Event queue and placed in the Output queue.