TRIGger
TRIGger
[:SEQuence]
:COUNt
N/A
1
The count specified in this command must be between 1 and 65535.
ABORt, ROUTe:SCAN
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence][:IMMediate]
TRIGger[:SEQuence]:DELay
This command specifies the number of times to sequence through an entire scan
sequence.
Command Response
route:scan
(@m1(1,2),m2(1:64))
Define a scan sequence consisting of relays 1 and 2 on relay
module m1 and relays 1 through 64 on module m2.
trigger:sequence:count
2
Define the number of times to sequence through the scan list
before setting the state of the trigger subsystem back to the
idle state.
trigger:sequence:
source:immediate
Sequence through the scan list without waiting for a trigger
event after an INITiate[:Immediate] command is received.
initiate:immediate Initiate the scan sequence. Sequence through the entire scan
list two times.