Triggering
Applies to sweep only. You can issue triggers for sweeps
using Internal (immediate) triggering, External triggering, or
Manual (software) triggering.
Internal triggering is enabled when you turn on the
U2761A. In this mode, the U2761A outputs continuously
when the sweep mode is selected.
External triggering uses the device Trig In connector to
control the sweep. The U2761A initiates one sweep each
time Trig In receives a TTL pulse. You can select whether
the U2761A is triggered on the rising or falling edge of
the External trigger signal.
Manual triggering initiates one sweep each time you click
Trigger on the soft front panel.
Trigger Source Choices
You must specify the source from which the U2761A accepts
a trigger.
The sweep trigger source consists of Internal, External, or
Manual. The default trigger source is Internal.
The U2761A accepts a Manual trigger, a hardware trigger
from the device Trig In connector, or continuously output
sweeps using an Internal trigger. At power-on, Internal
trigger is selected.
The trigger source setting is stored in volatile memory
where the source is set to Internal (immediate) trigger
upon power-off or after a remote interface reset.
Soft Front Panel Operation
On the Trigger Setup panel as shown in Figure 2-28, select
the desired trigger source from the drop down list
at the Source panel.