Agilent Technologies U2761A Computer Drive User Manual


 
90 U2761A User’s Guide
2 Features and Functions
Set Triggering Using SCPI Commands
Example 1
-> *CLS; *RST //To reset the U2761A to
default power-on state,
this command can be
ignored if this operation
is not required.
-> FUNC SIN //Sets the carrier waveform
to Sine wave.
-> VOLT 5 VPP //Sets the output amplitude
to 5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to
0.
-> SWE:STAT ON //Enables frequency sweep.
-> FREQ:STAR 500 //Changes the start
frequency to 500 Hz.
-> FREQ:STOP 10000 //Changes the stop
frequency to 10 kHz.
-> SWE:SPAC LIN //Changes the sweep mode
to Linear.
-> SWE:TIME 2 //Changes the sweep time
to 2 s.
-> TRIG:SOUR IMM //Changes the sweep trigger
source to Internal.
-> OUTP:TRIG ON //Turns on the trigger
output.
-> OUTP:TRIG:SLOP POS //Outputs the trigger on a
rising edge.
-> OUTP ON //Turns on output.