Agilent Technologies U2761A Computer Drive User Manual


 
Features and Functions 2
U2761A User’s Guide 89
When the Internal (immediate) trigger source is selected, the
U2761A outputs a Square waveform with a 50% duty cycle to
the Trig Out connector at the beginning of the sweep. The
period of the Square waveform is equal to the specified
sweep time.
When the Manual (software) trigger source is selected, the
U2761A outputs a pulse to the Trig Out connector at the
beginning of each sweep.
Soft Front Panel Operation
After enabling sweep, to specify whether the U2761A outputs
a TTL-compatible Square waveform with either a rising or
falling edge, select Positive for rising edge and
Negative for falling edge.
Remote Interface Operation
OUTPut:TRIGger {0|OFF|1|ON}
OUTPut:TRIGger:SLOPe {POSitive|NEGative}