Agilent Technologies U2761A Computer Drive User Manual


 
92 U2761A User’s Guide
2 Features and Functions
1 Select the Arbitrary waveform function.
Click Arb to select the Arbitrary function.
2 Set the waveform period.
Select Period and use the numeric keypad or
numeric drop down to set the period for the
waveform.
For example, set the period of the waveform to 10 ms.
3 Set the waveform voltage limits.
Select HiLevel and Lolevel to set the
high and low voltage levels that can be reached while
editing the waveform. The upper limit must be greater
than the lower limit.
For example, set the upper limit to 2.0 V and the lower
limit to 0 V.
4 Start the Arbitrary waveform editor.
Select Non Built-in Waveform from the Arbitrary’s drop down
list and click to start the Arbitrary
waveform editor.
5 Select the interpolation method.
Click to enable or disable linear interpolation between
the user-defined points. With interpolation enabled
(default), the waveform editor makes a straight-line
connection between the points. With interpolation
disabled, the waveform editor maintains a constant
voltage level between the points and creates a “step-like’