Barco R9003110 Projector User Manual


 
5. Getting started
Image 5-8
warning message in case of an MGP 10 projector
This warning message will be repeated at the next start up. Press BACK or MENU to remove the message.
The total lifetime of the lamp for a safe operation is 1500 hours max, do not use it longer. Always replace with a same type of lamp.
Call a BARCO authorized service technician for lamp replacement.
When the lamp runtime reaches 1500 hours the projector switches automatically to the other lamp, being lamp2.. following mes-
sages are displayed during and after switching.
Image 5-9
Image 5-10
When lamp2 at its turn
reaches 1470 hours, a warning message appears on the screen.
Image 5-11
warning message in case of a MGP 10 projector
At the end of the lifetime of lamp2 (1500 hours) the projector generates an alert message.
Image 5-12
A countdown time of 4 minutes is triggered before the projector is shut down (standby).
If the lamp runtime has not been reset, the alert message will reappear at the next start up (with again 4 minutes countdown time).
This al
ert message can be escaped with MENU or BACK, but the countdown continues.
Contact a qualified Barco technician for lamp replacement.
In Dual mode the lamp end of lifetimes are reached at the same time, however if in dual lamp mode one lamp
has been used more than the other (for example if the projector has been working temporarily in single mode),
one lamp will reach its end of lifetime sooner than the other lamp, which brings us to the Single mode oper-
ation.
5.6 Quick set up adjustments
Overview
Quick lens Adjustment
•UsingtheRCU
R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004
33