10. Settings
Full white
Clearbase
Bluebase
Video(6500K)
Computer(9300K)
Custom
Image 10-3
10.2 Display function
Overview
Display function in a standalone projector setup
Display function in a closed loop medical visualization system
10.2.1 Display function in a standalone projector setup
Display function setting
In a standalone or open loop system i.e. without any feedback from the screen, the display function is defined by the user via the
OSD of the projector. The selected display function corresponds then to a Look Up Table (LUT) as it is was set in the factory.
Factory settings are “dark room” settings meaning they do not take account of environmental conditions (am-
bient light, glare,...).
How to define a display function via the projector menu (standalone setup) ?
1. Press MENU to activate the Toolbar
2. Press to select the Settings item
3. Press to Pull down the Settings menu
4. Use or to select Display function
5. Press to pull down the menu
6. Use or to select the desired display function (image 10-4)
7. Press ENTER
The display function is adapted and the corresponding LUT is loaded.
R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004