10. Settings
Image 10-1
The custom color temperature allows the selection of a user defined color temperature set in the Adjustment
menu.
10.1.3 Color temperature in a closed loop medical visualization system
Color temperature in a closed loop medical visualization system
If, however the projector is part of a typical closed loop medical visualization system (Barco Dicom
®
Theatre ), color temperature
will be set by the control & calibration system (MediCal
®
Pro) along with a display function forming a pre-set value.
The preset (formed by color temperature and display function
) values are checked on a regular basis to verify if the projected image
is within a tolerance with respect to consistency, in the case it isn’t, a calibration process will be launched.
How to select a color temperature ?
1. Refer to the system manual
OSD interaction
WhenapresetisselectedonMediCal
®
Pro , the selected color temperature (as well as the display function) will be marked on the
projector’s OSD menu.
Full white
Clearbase
Bluebase
Video(6500K)
Computer(9300K)
Custom
Image 10-2
The completion of the calibration procedure will in the same way be signaled by a separate logo (as well as for the display function).
68