Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 93 www.PressureSystems.com
ii dd Description 'q' read
index
0C 00
01
Energize C2: Set Cal Valves to RUN or LEAK Position
(default) — choice made by ii=12.
Energize C1: Set Cal Valves to CAL/RE-ZERO or
PURGE Position — choice made by ii=12.
0B
see
chart
below
0D-0E -- Reserved for factory use
0F 00
01
Disable periodic Thermal Coefficient Update task.
Enable periodic Thermal Coefficient Update task
(default).
10 01-FF
(1-255)
Set Number of A/D Samples to Average. (default = 8). 05
11 -- Reserved for factory use
12 00
01
Energize C4: Set Cal Valves to RUN or CAL/RE-ZERO
Position (default) — choice made by ii=0C.
Energize C3: Set Cal Valves to PURGE or LEAK
Position — choice made by ii=0C.
0B
see
chart
below
13 00
01
Use Static IP Address Resolution (default)
Use Dynamic IP Address Resolution
06
14 00
01
02
Disable TCP/IP Back-Off Delay (default).
Enable TCP/IP Back-Off Delay as low-order byte of
Ethernet Adrs.(converted to decimal value * 20 µSec.).
Enable TCP/IP Back-Off Delay specified per eeee as
decimal value (* 20 µSec.)
07
15 -- Reserved for factory use
16 00
01
Disable TCP/IP Data Field Size Prefix (default).
Enable TCP/IP Data Field Size Prefix (2-byte big-endian
binary value with total size of response data in bytes).
08
17 00 Set TCP Connect Port per eeee as decimal value
(default=9000).
09
18 00
01
Disable Auto UDP Broadcast@Reset (default).
Enable Auto UDP Broadcast@Reset.
0A
19 00
01
Set Minimum Temperature Alarm Set Point (in
0
C)
per eeee as decimal value (default = 0
0
C).
Set Maximum Temperature Alarm Set Point ( in
0
C)
per eeee as decimal value (default = 60
0
C).
0D
0E
1A 00 Set External DH Scan Interval per eeee as decimal value
(seconds), 1 <= eeee <= 3600 (1 hour) (default = 60).
10