Pressure Systems 98RK Scanner User Manual


 
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 67 www.PressureSystems.com
Command 'c'— Sub-command Index 03: Clear Stream
This sub-command is used to “undefine” any previously configured host stream in a module,
one at a time, or all together. The sub-command’s format is:
Command
c 03 st
‘c’ is the command letter.
‘ 03’ is the sub-command index (ii) for Clear Stream.
st’ is the stream id digit (1, 2, or 3, or 0=all streams).
NOTE: all parameters are separated by a space.
ResponseA
‘A’ is the acknowledge letter.
Autonomous
Packet
none generated
Description: This sub-command clears (un-defines) the particular specified host stream (st=1-
3); or un-defines “all configured” host streams with a single command (st=0).
Once cleared, a stream must be reconfigured before it can be started.
Example:
Stop all the streams configured above. Then clear (undefine) only stream 3 and then
resume the remaining defined streams 1 and 2:
c 02 0
c 03 3
c 01 0
Read response:
"A"