Pressure Systems 98RK Scanner User Manual


 
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 48 www.PressureSystems.com
Example:
Supply each of the previously-specified three (3) pressure calibration points to the Multi-
Point Calibration function, as was stated in the previous example of the Configure and
Start (‘00’) sub-command. Assume that all the affected four (4) channels have
differential transducers with the same -5 to +5 psi range. Include at least one pressure
point in the negative range of these transducers.
C 01 1 -2.5
C 01 2 0.0
C 01 3 5.0
Read responses (separately after each command executed above):
"-2.4998 -2.4999 -2.5001 -2.500"
"0.0 0.0013 -0.0133 -0.00001"
"5.0091 4.9992 5.0010 4.9998"
Data are returned in reverse channel number order (i.e., 4, 3, 2, 1) in each response.