Pressure Systems 98RK Scanner User Manual


 
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 65 www.PressureSystems.com
For periodic hardware-triggered streams, that are never suspended and resumed
after being initially enabled, the sequence number may also serve as a “relative”
time stamp if the period (in milliseconds) of the hardware trigger is known.
If a special sub-command (ii=05) is used to select the content of a stream, other
status data may immediately follow the binary stream header and precede the
default Pressure EU Data (if selected). Still other special acquired data (per
selected channel) may follow the Pressure EU Data. The total of all groups of
selected acquired data in each packet will be ordered from highest channel
number requested to lowest channel number requested. Each such datum
(dddd) will be output per the format code specified when the stream was
configured (by sub-command 00).
Example:
Start all the streams configured in the previous example:
"c 01 0"
Read response:
"A"
Soon after the response is received, the requested data stream packets will begin
arriving in the host at a quantity, content, and rate determined by each stream’s own
particular current configuration (per both the 00 and 05 sub-commands).