Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 23 www.PressureSystems.com
Figure 2.5
Each 9816 is manufactured with a connection to one of the eight (8) CAL port ranges, based on
the unit’s full scale pressure range. Assignment of 9816's CAL range is show in Table 2.1 (next
page). In the case of a 9816 with multiple pressure ranges installed, the unit will use the CAL
range of the lowest installed full scale pressure range.
Note
CAL port assignments are based on a pressure range, NOT 98RK-1
slot assignments.
Table 2.1
Pneumatic Connector Assignment
98RK-1
Connection Pressure Scanner Range Input
Cal 1 User defined
Cal 2 10" W.C.; 20" W.C.; ±1 psid, ±1.5 psid
Cal 3 ±2.5; ±5 psid
Cal 4 ±10; ±15 psid, ±20 psid, ±25 psid
Cal 5 ±30; ±45; ±50 psid
Cal 6 ±75 psid, ±100 psid
Cal 7 +150; 200; 250; 300 psid
Cal 8 500 psid, 600 psid, 650 psid, 750 psid, 850
psid