Pressure Systems 98RK Scanner User Manual


 
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 55 www.PressureSystems.com
Note
The calculated gain values from the latest 'Z' command will be
lost when the module is powered OFF. To save these gain
terms to each transducer's nonvolatile memory, refer to the Set
Operating Options ('w') command (index 09).
Example:
Send TCP/IP command to a 9816 module (via its open socket) to calculate and set gain
coefficients for channels 8 through 4. Instruct the module to use 14.8890 psi as the
applied pressure instead of each transducer’s full-scale value:
Z00F8 14.8890
Response:
" 1.01289 1.06953 1.03750 0.99704 1.10051"
Actual gain values are returned in the above response as decimal ASCII strings, each
preceded by a space character. From left-to-right: they are for channels 8, 7, 6, 5, and
4.