Fluke 435 Power Supply User Manual


 
Tips and Maintenance
Troubleshooting22
22-3
Optional Accessories.
Item Ordering Code
Logger Functions for Fluke 434 (Mains Signaling, Logging). Fluke-434/Log
Optical Isolated RS-232 Cable PM9080
GPS Synchronization Unit GPS430
Print Adapter for Parallel Printers PAC91
Optical Isolated Trigger Probe (to test energy meters) ITP120
AC Current Clamp 200 A (10 mV/A) and 20 A (100 mV/A) switcheable. i200s
AC Current Clamp 2000 A (1 mV/A) and 200 A (10 mV/A) switcheable, flexible. i2000flex
AC Current Clamp 1000 A (1 mV/A), 100 A (10 mV/A), and 10 A (100 mV/A)
switcheable.
i1000s
AC Current Clamp 3000 A (0.1 mV/A), 300 A (1 mV/A), and 30 A (10 mV/A)
switcheable.
i3000s
AC/DC Current Clamp 100 A (10 mV/A) and 10 A (100 mV/A) switcheable. 80i-110s
AC Current Clamp 5 A (400 mV/A, 3 pack) i5s PQ3 (*)
Service Manual (English) www.fluke.com
(*): The SETUP/Clamp menu offers you a dedicated position to adapt the Analyzer for
use with i5s.
Troubleshooting
Analyzer does not start up.
The battery may be completely empty. In this case the Analyzer will not start up, even if
it is powered by the Battery Charger/Power Adapter. Charge the battery first: power the
Analyzer with the Battery Charger without turning it on. Wait about 15 minutes and try
turning on the Analyzer again.
Analyzer shuts down after a few seconds.
The battery may be empty. Check the battery symbol in the screen header. The
symbol indicates that the battery is empty and must be charged.
Attention: the Analyzer switches off automatically when powered by battery only if no
further knobs are operated after power-on (i.e. when the welcome screen is displayed).