Fluke 434/435
Users Manual
23-14
COMMON MODE REJECTION RATIO (CMRR)
CMRR >60 dB
SAFETY
Compliance
with
IEC/EN61010-1-2001,
CAN/CSA C22.2 No 61010-1-04 (including
c
CSA
us
approval),
UL std No 61010-1,
Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use, Part 1: General requirements,
Rated: 600V CAT IV 1000V CAT III Pollution Degree 2
Max voltage on
banana input
1000 V CAT III / 600 V CAT IV
Max voltage on
current BNC
input 42 Vpeak
ENVIRONMENTAL
Operating temperature 0°C to +50°C battery only, 0°C to +40°C with adapter, within spec +15°C to
+35°
Storage temperature -20 °C to +60 °C
Humidity 10 .. 30 °C: 95% RH non condensing
30 .. 40 °C: 75% RH non condensing
40 .. 50 °C: 45% RH non condensing battery only
Maximum operating
altitude
3000m. Derate to 1000 V CAT II / 600 V CAT III / 300 V CAT IV above 2000m
Maximum storage
altitude
12km
PRINTERS AND INTERFACE
Type Serial, optically isolated. Compatible with PM9080 (RS-232) or OC4USB
(USB)
Baud rate 1200, 2400, 9600 … 57k6
Print out facility (B&W
only)
Via optional adapter PM9080 or PAC 91
Print protocol Epson FX LQ, Deskjet, LaserJet , DPU-414 or PostScript