Quick Check® SV Series User’s Guide 6 - 31
Mode 01 (~LV01)
Mode 01 I/O operation
1. System will operate in Edge Sync or Envelope sync mode (~LX#)
2. Sync polarity programmable (~LP#)
3. Use ~PR0210 for proper port initialization.
4. Port 1 will go active OFF (no current) on an error condition.
5. Port 2 will go active ON (sink current) on an error condition.
6. Pushing the reset button or power re-cycle will place Ports 1 and 2 in their
inactive states and reset the # codes per sync counter.
7. Error conditions available are:
a. Partial Read (Programmable, ~LQ#, ~Lp#)
b. % decode (Programmable, ~LD##)
c. Bad Quiet Zone (~PB816xxx100)
d. ISO Defects grade (Programmable, ~PB806xxx000)
e. ISO Decodability (Programmable, ~PB802xxx100)
f. Symbol Contrast grade (Programmable, ~PB804xxx100)
g. No Read (if sync is received)
h. Number of codes per sync programmable (~LN##, ~LZ##)
i. Overall ISO Grade (Programmable, ~LAxx)
j. Symbology modulo check digit
Mode 01 LED Operation
1. LED1 will turn on if Outputs 1 and 2 go active due to ISO, contrast, or quiet
zone failure
2. LED2 will turn on if Outputs 1 and 2 go active due to a partial or no read
condition