5-18 A-61140 January 1999
5. Grasp and hold the new lower imaging guide by the green handle
and carefully slide it into place. Make sure the grooves fully engage
the track.
Rear imaging guide
saddle mount
Rear imaging guide
mounting pin
6. Hold the new front imaging guide by the edges and carefully set it
back into the scanner by seating the guide’s two green saddle
mounts on to the mounting shaft, and rotating the front imaging guide
thumbscrew 90° clockwise.
NOTE: The imaging guides are not interchangeable. They can only
be inserted one way.
Look through the upper access door and side access door
to make sure the imaging guide is mounted correctly. There
are two sets of pins. Make sure the guide is seated on the
small, lower set of pins. When the rear imaging guide is
seated correctly, the top of the guide will sit approximately a
½-inch (10 mm) from the vertical transport plate. The rear
imaging guide will become flush with the plate when the
plate is closed.
Front imaging
guide saddle mount
Vertical
transport plate