Triggering on Waveforms
3–96
TDS 500C, TDS 600B, & TDS 700C User Manual
Set the Mode and Holdoff. Mode and holdoff can be set for all standard trigger
types and classes. To set mode and holdoff, refer to Set Mode & Holdoff on
page 3–75. To learn more about trigger mode and holdoff, see Trigger Modes
and Trigger Holdoff on page 3–65.
When you select the pulse class Width, the oscilloscope will trigger on a pulse
narrower (or wider) than some specified range of time (defined by the upper
limit and lower limit). To set up for width triggering, do the following proce-
dures.
Select Width Triggering. Press TRIGGER MENU Type (main) Pulse
(pop-up) Class (main) Width (pop-up).
Select the Source. Press TRIGGER MENU Type (main) Pulse
(pop-up) Source (main) Ch1, Ch2, Ch3, or Ch4 (side). The source
selected becomes the trigger source for all four trigger classes.
Select the Polarity. Press TRIGGER MENU Type (main) Pulse
(pop-up) Class (main) Width (pop-up) Polarity (main) Positive or
Negative (side).
Set to Trig When. To set the range of widths (in units of time) the trigger source
will search for and to specify whether to trigger on pulses that are outside this
range or within this range, do the following steps:
1. Press TRIGGER MENU Type (main) Pulse (pop-up)
Class (main) Width (pop-up) Trig When (main).
2. Press Within Limits (side) if you want the oscilloscope to trigger on pulses
that fall within the specified range. If you want it to trigger on pulses that are
outside the range, then press Out of Limits (side).
Trigger Based on
Pulse Width