Chapter 7: Troubleshooting
To execute disaster recovery procedures
CD-ROM drive. The display will then show the boot devices:
Path Number Device Path Device Type
------------ ------------ ------------------
P0 SESCSI.6.0 IBM DNES-309170W
P1 SESCSI.1.0 PLEXTOR CD-ROM PX-40TS
7 At the prompt:
Main Menu: Enter command >
Type: BO P1 <Enter>
8 At the prompt:
Interact with IPL (Y, N, Q) ?>
Type: N <Enter>
9 After about 30 seconds you will see the message:
WARNING: The configuration information calls for a non-
interactive installation.
Press <Return> within 10 seconds to cancel batch mode
installation:
10 To abort the reinstallation process at this point:
Press the [Return] key on the keyboard within 10 seconds. (If you do nothing
within the 10 second time-out, the reinstallation process will begin. The
instrument will completely reload the operating system software onto the hard
disk drive.)
11 Processor Support Packages, Auxiliary Software, and user files must be
installed manually once the operating system has been reinstalled.
12 Follow the steps on page 15 to reload system settings and license
passwords.