Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
d Select the Channels field, select two channels to deskew, select
Run, and follow the instructions on the display.
e Repeat the channel skew procedure until all channel combinations
have been deskewed.
5 Remove the BNC cables from the instrument.
6 Select Close in the Calibration window.