Agilent Technologies 16700B Computer Accessories User Manual


 
82
Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
b Repeat for all cards in the module. Up to 4 cards may be configured
on a single time base and trigger in a 16700A, 16700B, 16702A, or
16702B mainframe
2 Reapply power to the 16700-series mainframe.
3 In the Logic Analysis System window select the icon for the master
card of the multi-card module then select Calibration.
4 Perform channel skew calibration on the multi-card module.
a Connect two 9-inch 50-ohm BNC cables and a BNC tee adapter
between channel 1, AC/DC cal, of the first card and channel 1 of the
second card.
Chan. 1
Chan. 1
Chan. 2
Chan. 2
Channel 1
AC/CD Cal
Channel 1
Upper Card
Lower Card
Upper Card