APPENDIX B: Batteries
Appendix B: Batteries
B.1 Batteries
OMNIScanner and OMNIRemote are powered by rechargeable, recyclable battery
packs.
Under various state and local laws, it may be illegal to dispose of this battery
pack in the municipal waste stream. Check with your local solid waste officials for
details in your area for recycling options or proper disposal.
To remove the battery pack, place the instrument face down and slide the
arrow-shaped tab downward. Then lift the battery pack up towards you. To install
the battery pack, insert the two tabs on the top of the battery in the Scanner and
press down on the battery pack until it clicks into place.
The batteries have a limited number of recharge cycles. Typically, capacity is
warranted by the battery manufacturer for about 200 recharge cycles. The capacity
of a battery is highly dependent on the way it is recharged. Generally, batteries that
are fully discharged before the recharge cycle may have a longer life than those
subject to extended cycles of shallow discharge. Conditions of shallow discharge
promote a phenomenon known as voltage depression, which gives the appearance
of reduced capacity. Battery life can also be affected by the temperature conditions
under which charging is performed. Charging at room temperature (77°F/25°C)
maximizes the battery life.
OMNIScanner and OMNIRemote use a charging mechanism that minimizes
the effects of shallow discharging. To maximize battery life, the battery
temperature is monitored and maintained within limits during the charging
process. Charging under high temperature conditions may result in longer
recharge times or premature termination of charging.