SCPI Command Reference
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the RF multiplexers. Commands are
listed alphabetically by subsystem and within each subsystem.
ABORt
The ABORt subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are
TRIGger:SOURce BUS or
TRIGger:SOURce HOLD.
Subsystem Syntax ABORt
Comments ABORt operation invalidates the current channel list and sets
ARM:COUNt 1 (one scanning cycle per INITiate command), sets
INITiate CONTinuous OFF (no continuous scanning cycles), and sets
TRIGger:SOURce IMMediate (continuous internal triggering).
Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface
CLEAR command or the
HP E1301A front panel “Reset Instr” or “Clear Instr” key to stop
the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS or TRIGger:SOURce HOLD, use ABORt or
the HP E1301A front panel “Reset Instr” or “Clear Instr” key to
stop the scan.
Stopping Scans Enabled from Front Panel: When a scan is
enabled from the HP E1301A front panel, execute
*RST entered via
the interface or the front panel “Reset Instr” or “Clear Instr” key to
stop the scan.
Related Commands: ARM, INITiate:CONTinuous, [ROUTe:]SCAN,
TRIGger
Example Stopping a Scan with ABORt
TRIG:SOUR BUS !
*TRG
command is trigger source
INIT:CONT ON ! Set continuous scanning
SCAN (@100:103) ! Scan channels 00-03
INIT ! Start scan, close channel 00
.
.
ABOR ! Abort scan in progress
40 HP E1366A/E1367A RF Multiplexers Command Reference Chapter 5