Agilent Technologies E1301B Computer Hardware User Manual


 
:PULSe:PERiod SOURce:PULSe:PERiod < period> sets the period of the pulse(s) to be
generated by the Pacer.
Parameters Parameter
Name
Parameter
Type
Range of
Values
Default
Units
pweiod numeric 500E-9 to 8.388607 or
MIN| MAX
second
Comments The resolution of period is 500E-9 seconds.
The Pacer waveform is a square wave with the output high for the first
half of the period, and low for the final half.
Related Commands: SOUR:PULS:COUN, ABORT, INIT,TRIG
*RST Condition: SOUR:PULS:PER 1E-6
Example Setting the Pacer pulse period.
TRIG:SOUR HOLD trigger source is TRIG
command
SOUR:PULS:COUN 1E3 output 1000 Pacer pulses
SOUR:PULS:PER .1 S pulse period set to .1 second
INIT go to Wait For Trigger state
TRIG trigger the Pacer to output
pulses
:PULSe:PERiod? SOURce:PULSe:PERiod? [MIN | MAX] returns :
The current period if no parameter is sent.
The maximum allowable period if MAX is sent.
The minimum allowable period if MIN is sent.
Example Querying the Pacer pulse period.
SOUR:PULS:PER? ask for pulse period
enter statement statement to enter value of
period
[SOURce] :PULSe:PERiod
System Instrument Command Reference 7-31