Agilent Technologies E1301B Computer Hardware User Manual


 
Switchbox Menu
Menu Levels and Content
Level 1 Level 2 Level 3 User Entry Command(s) Used Description
SWITCH MONITOR card number ‡ or AUTO DISP:MON:CARD < card_number> ;STAT ON Monitor instrument operations
OPEN channel list † OPEN (@channel_list) Open channel(s)
CLOSE channel list † CLOS (@channel_list) Close channel(s)
SCAN SET_UP channel list † TRIG:SOUR HOLD;:SCAN <channel_list>;:INIT Set up channels to scan
STEP channel list † TRIG Step to next channel in scan list
CARD TYPE? card number ‡ SYST:CTYP? < card_number> Display module ID information
DESCR? card number ‡ SYST:CDES? < card_number> Display module description
RESET card number ‡ SYST:CPON < card_number> Return module to power-on state
TEST * TST? Runs self-test, displays r esults ( + 0= pass; any
other number= fail)
† Channel lists are of the form “ccnn” (single channel), “ccnn,ccnn” (two or more channels) or “ccnn:ccnn” (range of channels); where “cc” is the card number and “nn” is the channel
number. For example, to access channel 2 on card number 1 specify 102.
‡ The card number identifies a module within the Switchbox. The switch module with the lowest logical address is always card number 01. The switch module with the next
successive logical address is card number 02 and so on.
2-16 Using the Display Terminal Interface