Error Messages and Causes
Code Message Cause
2110 Config error 10, Insufficient
system memory
Too many instruments installed for the amount
of RAM installed in the mainframe. Cannot
configure instruments. Only the system
instrument is started.
2111 Config error 11, Invalid
instrument address
A device’s logical address is not a multiple of 8
and the device is not part of a combined
instrument.
2113 Config error 13, Logical
address or IACK switch set
wrong
Duplicate logical addresses set or interrupt
bypass switches set improperly. Only the system
instrument is started.
2129 Config warning, Sysfail
detected
A device was asserting SYSFAIL on the
backplane during startup.
2130 Config error 30, Pseudo
instrument logical address
unavailable
A physical device has the same logical address as
IBASIC (240)
2131 Config error 32, File system
start up failed
Insufficient system resources to allow the
IBASIC file system to start.
2145 Config warning, Non-volatile
RAM contents lost
NVRAM was corrupted or a cold boot was
executed.
2148 Config warning, Driver RAM
contents lost
Driver RAM was corrupted or a cold boot was
executed.
2202 Unexpected interrupt from
non-message based card
A register based card interrupted when an
interrupt service routine had not been set up.
2809 Interrupt line has not been
set up
A DIAG:INT:ACT or DIAG:INT:RESP
command was executed before setting the
interrupt with DIAG:INT:SET.