Agilent Technologies E1301B Computer Hardware User Manual


 
TRIGger The TRIGger command subsystem controls the behavior of the trigger system
once it is initiated (see INITiate command subsystem). The trigger command
subsystem controls:
The delay between trigger and first Pacer pulse (TRIG:DELay)
An immediate software trigger (TRIG:IMM)
The source of the trigger (TRIG:SOUR BUS| EXT| HOLD| IMM)
Subsystem Syntax TRIGger
:DELay < delay>
:DELay? [MIN | MAX]
[:IMMediate]
:SLOPe < slope>
:SLOPe?
:SOURce BUS | EXT | HOLD | IMM
:SOURce?
:DELay TRIGger:DELay < delay> sets the delay between receipt of trigger and first
Pacer pulse.
Parameters Parameter
Name
Parameter
Type
Range of
Values
Default
Units
delay numeric 250E-9s to 4.19430375s or
MIN| MAX
second
Comments The resolution for delay is 250E-9 seconds.
Related Commands: ABORt, INITiate
*RST Condition: TRIG:DELay 2.5E-9
Example Setting delay between trigger and Pacer output.
TRIG:SOUR HOLD trigger is TRIG command
SOUR:PULS:COUN 100 set Pacer to output 100 pulses
SOUR:PULS:PER .1 S pulse period set to .1 second
TRIG:DELAY .75 S start Pacer .75 sec after trigger
INIT go to Wait For Trigger state
TRIG trigger Pacer to output pulses
:DELay? TRIGger:DELay? [MIN | MAX] returns:
The current delay if no parameter is sent.
The maximum allowable delay if MAX is sent.
The minimum allowable delay if MIN is sent.
Example Querying the trigger delay setting.
TRIG:DEL .75 S start Pacer .75 sec after trigger
TRIG:DEL? command System Instrument
to send TRIG:DEL value.
enter statement input value of trigger delay
TRIGger :DELay
System Instrument Command Reference 7-51