Agilent Technologies Z5623A Network Router User Manual


 
4-4 Agilent Technologies Z5623A Option H48
Calibrating the Network Analyzer
Performing the Calibration
Figure 4-4 Cal Type Window When It Is First Opened
Step 6. Click the Full SOLT 2-Port button. The result is shown in Figure 4-5.
Figure 4-5 Cal Type Window After Selecting “Full SOLT 2-Port”
Step 7. Click the OK button. The Cal Type window will then close and the Cal
Wizard window will pop back up as shown in Figure 4-6.
Figure 4-6 Cal Wizard Menu After Selection Of Cal Type Is Complete
Step 8. Verify that the currently selected calibration type shown in the Cal
Wizard window is “Full SOLT 2-Port (Omit Isolation)” as shown in
Figure 4-6.
Step 9. Click the Measure ECal Stds button in the Cal Wizard window. The
window shown in Figure 4-7 will then pop up.