Agilent Technologies Z5623A Option H48 4-5
Calibrating the Network Analyzer
Performing the Calibration
Figure 4-7 Electronic Calibration Window
Step 10. Click the Module button. The window shown in Figure 4-8 will then
pop up.
Figure 4-8 Module Connection Message
Step 11. Click the OK button. (The connections requested in Figure 4-8 were
made in Step 1 of “Performing the Calibration” on page 4-2.) This
window will close, the window shown in Figure 4-7 will pop back up,
and the calibration will begin.
The network analyzer will complete the calibration with no further
operator actions required. The calibration will take about 5 seconds.
Step 12. Verify that the calibration has been completed. This is shown by a check
mark over the Module button as shown in Figure 4-9.
Figure 4-9 Electronic Calibration Window After Cal Is Complete
Step 13. Click the OK button shown in Figure 4-9. The window will close; the
Cal Wizard window will pop back up with the following message as
shown in Figure 4-10: “Done. Save your settings now by pressing OK”.