Agilent Technologies Z5623A Network Router User Manual


 
Agilent Technologies Z5623A Option H48 5 -5
Performance Verification
Test Strategy
Insertion Loss
For this measurement, Port 1 of the network analyzer will be connected
to one of the test ports of the test set, and Port 2 of the network
analyzer will be connected either to the Reflection port or to the
Transmission port of the test set. By definition,
Insertion Loss = S
21
where Insertion Loss and S
21
are both in units of dB.
The network analyzer will be configured to measure S
21
. Limit lines
will be set up in such a way that each limit equals the negative of the
corresponding insertion loss test limit shown in Table 5-2. For example,
if an insertion loss test limit happens to be 1.0 dB, the corresponding
S
21
limit will be 1.0 dB.
Insertion loss is a maximum spec. Therefore, each S
21
limit will be set
up as a minimum limit. For example, if an insertion loss test limit
happens to be 1.0 dB, the limit table will be set up so that S
21
must be
above 1.0 dB in order for the instrument to pass. (S
21
will always be
zero or negative because the test set provides no amplification. In units
of dB, S
21
can range from zero to minus infinity.)
Return Loss
For this measurement, Port 1 of the network analyzer will be connected
to one of the test ports of the test set, and Port 2 of the network
analyzer will be connected to the Reflection port of the test set. By
definition,
Return Loss = S
11
where Return Loss and S
11
are both in units of dB.
The network analyzer will be configured to measure S
11
. Limit lines
will be set up in such a way that each limit equals the negative of the
corresponding return loss test limit shown in Table 5-2. For example, if
a return loss test limit happens to be 20 dB, the corresponding S
11
limit
will be 20 dB.
Return loss is a minimum spec. Therefore, each S
11
limit will be set up
as a maximum limit. For example, if a return loss test limit happens to
be 20 dB, the limit table will be set up so that S
11
must be below 20 dB
in order for the instrument to pass. (S
11
will always be zero or negative
because the test set provides no amplification. In units of dB, S
11
can
range from zero to minus infinity.)