Fluke 2625A Power Supply User Manual


 
Introduction and Specifications
Specifications
1
1-21
Table 1-4. 2635A Specifications (cont)
Input Impedance
100 M minimum in parallel with 150 pF maximum for all ranges 3V and below 10 M in parallel
with 100 pF maximum for the 30V and 300V ranges.
Normal Mode Rejection
53 dB minimum at 60 Hz ±0.1%, slow rate
47 dB minimum at 50 Hz ±0.1%, slow rate
Common Mode Rejection
120 dB minimum at dc, 1 k imbalance, slow rate
120 dB minimum at 50 or 60 Hz ±0.1%, 1 k imbalance, slow rate
Maximum Input
300V dc or ac rms on any range for channels 0, 1, and 11
150V dc or ac rms for channels 2 to 10 and 12 to 20
Voltage ratings between channels must not be exceeded
Crosstalk Rejection
Refer to "Crosstalk Rejection" at the end of Table 1-3.