National Instruments VXI-MIO Series Switch User Manual


 
Chapter 3 Hardware Overview
National Instruments Corporation 3-13 VXI-MIO Series User Manual
Figure 3-5.
Above-High-Level Analog Triggering Mode
In inside-region analog triggering mode, the trigger is generated when
the signal value is between the
lowValue
and the
highValue.
Figure 3-6.
Inside-Region Analog Triggering Mode
In high-hysteresis analog triggering mode, the trigger is generated
when the signal value is greater than highValue, with the hysteresis
specified by lowValue.
Figure 3-7. High-Hysteresis Analog Triggering Mode
highValue
Trigger
highValue
Trigger
lowValue
highValue
Trigger
lowValue