National Instruments VXI-MIO Series Switch User Manual


 
Appendix A Specifications for VXI-MIO-64E-1
VXI-MIO Series User Manual A-2
National Instruments Corporation
Input coupling....................................DC
Max working voltage
(signal + common mode).................Each input should remain within
±
11 V of ground
Overvoltage protection.......................
±
25 V powered on,
±
15 V
powered off
Inputs protected.......................... ACH<0..63>, AISENSE,
AISENSE2
FIFO buffer size.................................8,192 S
Data transfers.....................................DMA, interrupts, programmed
I/O
Configuration memory size................512 words
Input signal ranges ................
Module Gain
(Software
Selectable)
Module Range
(Software Selectable)
Bipolar Unipolar
0.5
±
10 V
1
±
5 V
0 to 10 V
2
±
2.5 V
0 to 5 V
5
±
1 V
0 to 2 V
10
±
500 mV
0 to 1 V
20
±
250 mV
0 to 500 mV
50
±
100 mV
0 to 200 mV
100
±
50 mV
0 to 100 mV