4-4 CONTROL FUNCTIONS
MON2000
Calibration JULY 2010
4.4 CALIBRATION
Use this function to calibrate the GC unit.
Calibration runs are determined by the CDT
and Streams settings. See Section 5.2 and
Section 5.8 for detailed instructions on how to
edit these settings.
1. Use the Control > Calibration menu to
access this function.
2. The Start Calibration dialog appears.
MON2000 displays the stream to be used
for the calibration.
To select which stream is used for
calibration, See Section 5.8.
3. Click the Purge check box to set the purging
option. Note that the default setting is ON.
If the GC unit is in Auto Sequence mode,
calibration will not start until 2 or more
analysis runs have been completed. This delay
is required to complete the current analysis
and the analysis of the stream currently
purging through the valve.
See Section 7.1 to select
report parameters.